- PMP9771, Energy Storage Circuit for Uninterrupted Power Supply and Extending Alkaline Battery Lifetime. An alkaline battery and a boost converter are usually used to power the load system which requires high voltage, such as the motor driver in insulin pump. In this reference design, a lithium polymer battery is added to the output of the boost converter to absorb the pulse load current and extend the alkaline battery lifetime. The designed circuit also benefits uninterrupted power supply when the alkaline battery is out of charge
Caractéristiques principales
-
Conversion TypeDC to DC
-
Input Voltage0.7 to 3.2 V
-
Output Voltage4 V
-
Output Current0.15 A
-
Output Power0.6 W
-
Efficiency70 %
-
TopologyBoost- Synchronous
Pièces en vedette (2)
Pièce numéro | Fabricant | Type | Description | |||
---|---|---|---|---|---|---|
|
TPS61220DCKR | Texas Instruments | DC to DC Converter and Switching Regulator Chip | Conv DC-DC 0.7V to 5.5V Synchronous Step Up Single-Out 1.8V to 5.5V 0.5A 6-Pin SC-70 T/R | Ajouter au panier | |
|
TPS61220DCKT | Texas Instruments | DC to DC Converter and Switching Regulator Chip | Conv DC-DC 0.7V to 5.5V Synchronous Step Up Single-Out 1.8V to 5.5V 0.5A 6-Pin SC-70 T/R | Ajouter au panier |